Introduction
1
Keysight U8030A Series User’s Guide
23
5
Adjust the voltage limit value to 1.00 V.
Press the
[Display Limit]
button to set the display to the limit mode (the
LIMIT
annunciator turns on).
Adjust the voltage limit to 1.00 V to ensure constant current (CC) operation.
The
CC
annunciator will turn on.
Press the
[Display Limit]
button again or allow the power supply to idle for a
few seconds to exit the limit mode.
6
Check that the front-panel ammeter properly responds to the knob control for
Output 1.
Press the
[Voltage/Current]
button. Ensure that the
A
annunciator is blinking
before proceeding with the next step.
Turn the knob clockwise or counterclockwise when the display is in the meter
mode (the
LIMIT
annunciator is off). Check that the ammeter responds to the
knob control and the voltmeter indicates nearly zero (the voltmeter will show
the voltage drop caused by the test lead).
7
Ensure that the current can be adjusted from zero to the full rated value by
adjusting the knob.
8
Repeat
for Output 2. Press
[1]
or
[2]
to switch between Output
1 and Output 2 (the
OUT1
and
OUT2
annunciators will turn on and off
respectively).
9
Disable the outputs, turn off the power supply, and remove the short from the
output terminals.
Summary of Contents for U8030A Series
Page 2: ...Keysight U8030A Series Triple Output DC Power Supply User s Guide ...
Page 9: ...8 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 13: ...12 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 15: ...14 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 17: ...16 Keysight U8030A Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...