1
Calibration Procedures
30
Keysight U1270 Series Service Guide
6
DCV
Turn the rotary switch to
the /
position.
3 V
3 V
±0.0020 V
±0.0020 V
30 V
30 V
±0.017 V
±0.017 V
300 V
300 V
±0.17 V
±0.17 V
1000 V
1000 V
±0.7 V
±0.7 V
7
AC+DCV
Turn the rotary switch
to the position, and press the
key twice.
3 V
3 V, 20 Hz
-
±0.0360 V
3 V, 45 Hz
-
±0.0235 V
3 V, 65 Hz
-
±0.0235 V
3 V, 1 kHz
-
±0.0360 V
3 V, 5 kHz
-
±0.0510 V
3 V, 20 kHz
-
±0.0675 V
2.7 V, 100 kHz
-
±0.1017 V
30 V
30 V, 20 Hz
-
±0.360 V
30 V, 45 Hz
-
±0.235 V
30 V, 65 Hz
-
±0.235 V
30 V, 1 kHz
-
±0.360 V
30 V, 5 kHz
-
±0.510 V
30 V, 20 kHz
-
±0.675 V
27 V, 100 kHz
-
±1.017 V
300 V
300 V, 45 Hz
-
±2.35 V
300 V, 65 Hz
-
±2.35 V
300 V, 1 kHz
-
±3.60 V
300 V, 5 kHz
-
±5.10 V
270 V, 20 kHz
-
±6.12 V
Table 1-2
Performance verification tests (continued)
Step Test function
Range
5520 output
Error from nominal 1 year
U1271A
U1272A/
U1273A/U1273AX
Shift
View
Esc
Summary of Contents for U1271A