Calibration Procedures
1
Keysight U1270 Series Service Guide
23
The alert beep tone will continue to beep until the test lead is removed from the
A
terminal or until the rotary switch is set to the
function.
Input warning test (µA mA terminal)
The multimeter sounds a continuous alert beep when the test lead is inserted into
the
µA mA
terminal but the rotary switch is not set to the or
function.
The multimeter displays an input warning error (
).
Figure 1-4
Input warning display (µA mA terminal)
The alert beep tone will continue to beep until the test lead is removed from the
µA mA
terminal or until the rotary switch is set to the or
function.
NOTE
Before conducting this test, ensure that the beep function is not disabled in the
multimeter’s setup.
NOTE
Before conducting this test, ensure that the beep function is not disabled in the
multimeter’s setup.