Calibration Procedures
1
Keysight U1270 Series Service Guide
19
Calibration interval
A 1-year interval is adequate for most applications. Accuracy specifications are
warranted only if adjustment is made at regular calibration intervals. Accuracy
specifications are not warranted beyond the 1-year calibration interval. Keysight
does not recommend extending calibration intervals beyond 2 years for any
application.
Adjustment is recommended
Specifications are only guaranteed within the period stated from the last
adjustment. Keysight recommends that re-adjustment should be performed
during the calibration process for best performance. This will ensure that the
instrument will remain within the specifications for the next calibration interval.
This criterion for the re-adjustment provides the best long-term stability.
Performance data are measured during the
“Performance Verification Tests”
but
this does not guarantee that the instrument will remain within these limits unless
the adjustments are performed.
Refer to the
on page 47 and verify that all the adjustments
have been performed.