1
Calibration Procedures
32
Keysight U1270 Series Service Guide
10
Resistance
Turn the rotary
switch to the
/ position.
30
Ω
30
Ω
-
±0.070
Ω
300
Ω
300
Ω
±0.65
Ω
±0.65
Ω
3 k
3 k
Ω
±0.0065 k
Ω
±0.0065 k
Ω
30 k
30 k
Ω
±0.065 k
Ω
±0.065 k
Ω
300 k
Ω
300 k
Ω
±0.65 k
Ω
±0.65 k
Ω
3 M
Ω
3 M
Ω
±0.0185 M
Ω
±0.0185 M
Ω
30 M
30 M
Ω
±0.365 M
Ω
±0.365 M
Ω
100 M
100 M
Ω
±2.10 M
Ω
-
300 M
120 M
Ω
-
±9.70 M
Ω
[c] The accuracy of the 30
Ω
to 3 k
Ω
range is specified after the Null function is used to subtract the test lead resistance and thermal effect
(by shorting the test leads). Apply a 0
Ω
calibrator output and allow the multimeter to settle before pressing the
key.
[d] The RH is specified for <60%.
[e] With a 2-wire connection and compensation enabled at calibrator.
11
Diode
Turn the rotary switch to
the
or position.
3 V
3 V
0.0155 V
0.0155 V
12
Capacitance
Turn the rotary
switch to the position.
10 nF
10 nF
±0.105 nF
±0.105 nF
100 nF
100 nF
±1.02 nF
±1.02 nF
1000 nF
1000 nF
±10.2 nF
±10.2 nF
10
μ
F
10
μ
F
±0.102
μ
F
±0.102
μ
F
100
μ
F
100
μ
F
±1.02
μ
F
±1.02
μ
F
1000
μ
F
1000
μ
F
±10.2
μ
F
±10.2
μ
F
10 mF
10 mF
±0.102 mF
±0.102 mF
[f] The accuracy for all ranges is specified based on a film capacitor or better, and after the Null function is used to subtract the residual values
(by opening the test leads).
Table 1-2
Performance verification tests (continued)
Step Test function
Range
5520 output
Error from nominal 1 year
U1271A
U1272A/
U1273A/U1273AX
Smart
Scale
Null
Auto
Summary of Contents for U1271A
Page 1: ...Keysight U1270 Series Handheld Digital Multimeters Service Guide ...
Page 10: ...10 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...