Calibration Procedures
1
Keysight U1270 Series Service Guide
43
ACmA/ACA
30 mA
03.000 mA (70 Hz)
0.9 to 1.1
×
Reference value
30.000 mA (70 Hz)
0.9 to 1.1
×
Reference value
300 mA
030.00 mA (70 Hz)
0.9 to 1.1
×
Reference value
300.00 mA (70 Hz)
0.9 to 1.1
×
Reference value
3 A
0.3000 A (70 Hz)
0.9 to 1.1
×
Reference value
3.0000 A (70 Hz)
0.9 to 1.1
×
Reference value
10 A
3.0000 A (70 Hz)
0.9 to 1.1
×
Reference value
10.000 A (70 Hz)
0.9 to 1.1
×
Reference value
Capacitance
OPEN
OPEN
With cables connected to CAP/COM terminals
10 nF
04.000 nF
0.9 to 1.1
×
Reference value
10.000 nF
0.9 to 1.1
×
Reference value
100 nF
010.00 nF
0.9 to 1.1
×
Reference value
100.00 nF
0.9 to 1.1
×
Reference value
1000 nF
0100.0 nF
0.9 to 1.1
×
Reference value
1000.0 nF
0.9 to 1.1
×
Reference value
10
μ
F
10.000
μ
F
0.9 to 1.1
×
Reference value
100
μ
F
100.00
μ
F
0.9 to 1.1
×
Reference value
1000
μ
F
1000.0
μ
F
0.9 to 1.1
×
Reference value
10 mF
10.000 mF
0.9 to 1.1
×
Reference value
OPEN
OPEN
OPEN terminals (no cables connected to CAP/
COM terminals)
[a]
Table 1-3
Adjustment input values (continued)
Test function
Step
Reference value
Valid reference input
Summary of Contents for U1271A
Page 1: ...Keysight U1270 Series Handheld Digital Multimeters Service Guide ...
Page 10: ...10 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...