1
Calibration Procedures
20
Keysight U1270 Series Service Guide
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures are listed below (
). If the exact instrument is not
available, substitute with another calibration standard of equivalent accuracy.
A suggested alternative method is to use the Keysight 3458A 8½ Digit Digital
Multimeter to measure less accurate but stable sources. The output value
measured from the source can be entered into the instrument as the target
calibration value.
Table 1-1
Recommended test equipment
Application
Recommended equipment
Recommended accuracy requirements
DC voltage
Fluke 5520A
<20% of the U1270 Series accuracy specification
DC current
Fluke 5520A
<20% of the U1270 Series accuracy specification
Resistance
Fluke 5520A
<20% of the U1270 Series accuracy specification
AC voltage
Fluke 5520A
<20% of the U1270 Series accuracy specification
AC current
Fluke 5520A
<20% of the U1270 Series accuracy specification
Frequency
Fluke 5520A
<20% of the U1270 Series accuracy specification
Capacitance
Fluke 5520A
<20% of the U1270 Series accuracy specification
Duty cycle
Fluke 5520A
<20% of the U1270 Series accuracy specification
Diode
Fluke 5520A
<20% of the U1270 Series accuracy specification
Temperature
Fluke 5520A
<20% of the U1270 Series accuracy specification
TM Electronics KMPC1MP
(K-Type thermocouple extension)
—
Short
Pomona MDP-S
—
Z
low
Keysight U1252B
—
Summary of Contents for U1271A
Page 1: ...Keysight U1270 Series Handheld Digital Multimeters Service Guide ...
Page 10: ...10 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...