Calibration Procedures
1
Keysight U1270 Series Service Guide
41
Valid adjustment input values
Adjustment can be accomplished using the following input values below.
Table 1-3
Adjustment input values
Test function
Step
Reference value
Valid reference input
DCmV
SHORT
SHORT
SHORT V/COM terminals
30 mV
30.000 mV
0.9 to 1.1
×
Reference value
300 mV
300.00 mV
0.9 to 1.1
×
Reference value
ACmV
30 mV
3.000 mV (70 Hz)
0.9 to 1.1
×
Reference value
30.000 mV (70 Hz)
0.9 to 1.1
×
Reference value
30.000 mV (30 kHz)
0.9 to 1.1
×
Reference value
300 mV
30.00 mV (70 Hz)
0.9 to 1.1
×
Reference value
300.00 mV (70 Hz)
0.9 to 1.1
×
Reference value
300.00 mV (30 kHz)
0.9 to 1.1
×
Reference value
DCV
SHORT
SHORT
SHORT V/COM terminals
3 V
3.0000 V
0.9 to 1.1
×
Reference value
30 V
30.000 V
0.9 to 1.1
×
Reference value
300 V
300.00 V
0.9 to 1.1
×
Reference value
1000 V
1000.0 V
0.9 to 1.1
×
Reference value