Calibration Procedures
42
Keysight U1270 Series Service Guide
ACV
3 V
0.3000 V (70 Hz)
0.9 to 1.1
×
Reference value
3.0000 V (70 Hz)
0.9 to 1.1
×
Reference value
3.0000 V(3 kHz)
0.9 to 1.1
×
Reference value
30 V
3.000 V (70 Hz)
0.9 to 1.1
×
Reference value
30.000 V (70 Hz)
0.9 to 1.1
×
Reference value
30.000 V(3 kHz)
0.9 to 1.1
×
Reference value
300 V
30.00 V (70 Hz)
0.9 to 1.1
×
Reference value
300.00 V (70 Hz)
0.9 to 1.1
×
Reference value
300.00 V (3 kHz)
0.9 to 1.1
×
Reference value
1000 V
30.0 V (70 Hz)
0.9 to 1.1
×
Reference value
300.0 V (70 Hz)
0.9 to 1.1
×
Reference value
300.0 V (3 kHz)
0.9 to 1.1
×
Reference value
DC
μ
A
OPEN
OPEN
OPEN terminals
300
μ
A
300.00
μ
A
0.9 to 1.1
×
Reference value
3000
μ
A
3000.0
μ
A
0.9 to 1.1
×
Reference value
AC
μ
A
300
μ
A
030.00
μ
A (70 Hz)
0.9 to 1.1
×
Reference value
300.00
μ
A (70 Hz)
0.9 to 1.1
×
Reference value
3000
μ
A
300.0
μ
A (70 Hz)
0.9 to 1.1
×
Reference value
3000.0
μ
A (70 Hz)
0.9 to 1.1
×
Reference value
DCmA/DCA
OPEN
OPEN
OPEN terminals
30 mA
30.000 mA
0.9 to 1.1
×
Reference value
300 mA
300.00 mA
0.9 to 1.1
×
Reference value
3 A
3.0000 A
0.9 to 1.1
×
Reference value
10 A
10.000 A
0.9 to 1.1
×
Reference value
Table 1-3
Adjustment input values (continued)
Test function
Step
Reference value
Valid reference input
Summary of Contents for U1271A
Page 1: ...Keysight U1270 Series Handheld Digital Multimeters Service Guide ...
Page 10: ...10 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...