1
Calibration Procedures
44
Keysight U1270 Series Service Guide
Resistance
SHORT
0
Ω
[b]
0
Ω
from calibrator
30 M
Ω
OPEN
OPEN terminals
10.000 M
Ω
0.9 to 1.1
×
Reference value
3 M
Ω
3.0000 M
Ω
0.9 to 1.1
×
Reference value
300 k
Ω
300.00 k
Ω
0.9 to 1.1
×
Reference value
30 k
Ω
30.000 k
0.9 to 1.1
×
Reference value
3 k
Ω
3.0000 k
0.9 to 1.1
×
Reference value
300
Ω
300.00
0.9 to 1.1
×
Reference value
30
Ω
30.000
0.9 to 1.1
×
Reference value
SHORT
SHORT
[c]
SHORT
Ω
/COM terminals
Temperature
K type
0000.0 °C
0 °C with ambient compensation required
Note:
Ensure the multimeter is turned on and stabilized for at least 60 minutes with the K-type thermocouple connected
between the multimeter and the calibrator output terminal.
– Set the 5520A to internal reference.
– Prior to performing adjustment, connect one end of the K-type thermocouple (with miniature TC connector on both ends)
to the 5520A TC output, and the other end to a precision thermometer to verify that the source outputs the desired value.
Adjust the source accordingly if necessary.
– To perform the adjustment, connect one end of the K-type thermocouple (with miniature TC connector on both ends) to the
5520A TC output, and the other end to the multimeter via a TC-to-banana adapter. Allow at least 1 hour for the multimeter
to stabilize.
Diode
SHORT
SHORT
SHORT V/COM terminals
3 V
2.0000 V
0.9 to 1.1
×
Reference value
[a] After capacitance calibration is completed, calibrate the OPEN item again by removing all cable connections from the Device-Under-Test (DUT).
[b] With a 2-wire connection and compensation enabled at the calibrator.
[c] After resistance calibration is completed, calibrate the SHORT item again by inserting a short bar between the
Ω
/COM terminals.
Table 1-3
Adjustment input values (continued)
Test function
Step
Reference value
Valid reference input
Summary of Contents for U1271A
Page 1: ...Keysight U1270 Series Handheld Digital Multimeters Service Guide ...
Page 10: ...10 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight U1270 Series Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...