Service
2
Keysight E3632A Service Guide
69
631
System DAC test failed
This test checks if the DAC hardware is functional. The main controller U17 (
U10 for serial MY53xx6xxx
)
sends a reference voltage data to DAC and converts the DAC output to digital data to see if the digital data
is within a valid range.
632
Hardware test failed
This test checks the status of voltage and current error amplifiers for the power circuit of output1. If both
amplifiers are not operational, the power supply will beep and the error annunciator will be lit on.