1
Calibration Procedures
38
Keysight E3632A Service Guide
CC performance test record
Table 1-3
CC performance test record
Test description
Actual
result
Specifications
Upper limit
Lower limit
CC programming accuracy @ 0 A (I
O
)
+0.0100 A
–0.0100 A
CC readback accuracy @ 0 A
I
O
+ 0.0050 A
I
O
– 0.0050 A
CC programming accuracy @ full scale
(I
O
)
7.0240 A
6.9760 A
CC readback accuracy @ full scale
I
O
+ 0.0155 A
I
O
– 0.0155 A
CC load regulation
Maximum change: <0.95 mA
CC line regulation
Maximum change: <0.95 mA
CC ripple/noise
<2 mA
Common mode current noise
<1.5
μ
Arms