Load transient response time
This test measures the time for the output voltage to recover to within 15 mV of
nominal output voltage following a load change from full load to half load, or half
load to full load.
1
Turn off the power supply and connect the output to be tested as shown in
with an oscilloscope. Operate the electronic load in constant
current mode.
2
Turn on the power supply. Select the 30 V/4 A range, enable the outputs, and
set the display to the limit mode. When the display is in the limit mode,
program the current to the full rated value 4.0 A and the voltage to the full
rated value (30.0 V).
3
Set the electronic load to transient operation mode between one half of the
output's full scale value and the output's full rated value at a 1 kHz rate with
50% duty cycle.
4
Set the the oscilloscope for AC coupling, internal sync, and lock on either the
positive or negative load transient.
5
Adjust the the oscilloscope to display transients as shown in
that the pulse width (t2–t1) of the transients at 15 mV from the base line is no
more than 50 msec for the output.
Figure 1-2
Transient response time
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply Service Guide ...
Page 12: ...Keysight E3631A Service Guide 14 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...16 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 68: ...2 Service 70 Keysight E3632A Service Guide ...
Page 70: ...72 Keysight E3632A Service Guide Component locator diagram for the front panel top ...