Calibration Procedures
1
Keysight E3632A Service Guide
27
Constant Voltage (CV) Verifications
Constant voltage test setup
If more than one meter or a meter and an oscilloscope are used, connect each to
the (+) and (–) terminals by a separate pair of leads to avoid mutual coupling
effects. Use a coaxial cable or shielded 2-wire cable to avoid noise pick-up on the
test leads.
Voltage programming and readback accuracy
This test verifies that the voltage programming and the GPIB or RS-232 readback
functions are within specifications. Note that the readback values over the remote
interface should be identical to those displayed on the front panel.
1
Turn off the power supply and connect a digital voltmeter between the (+) and
(–) terminals of the output to be tested as shown in
.
2
Turn on the power supply. Select the 30 V/4 A range and enable the outputs
by sending the commands:
VOLT:RANG P30V
OUTP ON
3
Program the output voltage to zero volt and current to full rated value (4.0 A)
by sending the commands:
VOLT 0
CURR 4
4
Record the output voltage reading on the digital voltmeter (DVM). The reading
should be within the limits of 0 V ±10 mV. Note that the CV, Adrs, Limit, and
Rmt annunciators are on.
5
Readback the output voltage over the remote interface by sending the
command:
MEAS:VOLT?
NOTE
You should program the power supply over the remote interface for this test to
avoid round-off errors.
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply Service Guide ...
Page 12: ...Keysight E3631A Service Guide 14 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...16 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 68: ...2 Service 70 Keysight E3632A Service Guide ...
Page 70: ...72 Keysight E3632A Service Guide Component locator diagram for the front panel top ...