Calibration Procedures
1
Keysight E3632A Service Guide
37
Performance Test Record for E3632A
CV performance test record
Table 1-2
CV performance test record
Test description
Actual
result
Specifications
Upper limit
Lower limit
CV programming accuracy @ 0 volts
(DVM reading)
+0.0100 V
–0.0100 V
CV readback accuracy @ 0 volts
DVM + 0.0050 V
DVM – 0.0050 V
CV programming accuracy @ full scale
(DVM reading)
+30.025 V
29.9750 V
CV readback accuracy @ full scale
DVM + 0.0200 V
DVM – 0.0200 V
CV load regulation
Maximum change: <5
mV
CV line regulation
Maximum change: <5
mV
CV ripple/noise
<2 mVp-p, 0.35 mVrms
Load transient response time
<50
μ
sec