Calibration Procedures
1
Keysight E3632A Service Guide
47
I MI 3.5000 A
If the entered number is within an acceptable range, an ENTERED message
appears for one second. If the entered number is not correct, a MIN VALUE or
MAX VALUE message appears for one second and the display shows the low
current calibration point again. The display now shows the middle current
calibration point.
16
Read the DVM and change the middle current value on the display to match
the computed current (DVM reading divided by shunt resistance).
For example, if the computed value is 3.499 A, adjust the current to 3.499 A
using the knob and arrow keys.
I MI 3.4990 A
17
Pressing the Calibrate key saves the change and selects the high current
calibration point.
I HI 6.9000 A
If the entered number is within an acceptable range, an ENTERED message
appears for one second. If the entered number is not correct, a MIN VALUE or
MAX VALUE message appears for one second and the display shows the
middle current calibration point again. The display now shows the high current
calibration point.
18
Read the DVM and change the high current value on the display to match the
computed current (DVM reading divided by shunt resistance).
For example, if the computed value is 6.899 A, adjust the current to 6.899 A
using the knob and arrow keys.
I HI 6.8990 A
NOTE
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
NOTE
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply Service Guide ...
Page 12: ...Keysight E3631A Service Guide 14 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...16 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 68: ...2 Service 70 Keysight E3632A Service Guide ...
Page 70: ...72 Keysight E3632A Service Guide Component locator diagram for the front panel top ...