1
Calibration Procedures
48
Keysight E3632A Service Guide
19
Pressing the Calibrate key saves the new calibration constants for the output
current and goes to the OCP calibration mode.
CAL SETUP 4
A CALIBRATING message appears for one second to indicate that the current
calibration is progressing and new calibration constants of “SETUP 3” are
stored. Then, the display shows the above message to indicate that the power
supply is ready for the OCP calibration.
If the calibration fails, a DAC CAL FAIL or ADC CAL FAIL message appears for
one second and the display shows the CAL SETUP 3 for current calibration
again.
20
Pressing the Calibrate key saves the new OCP calibration constants and return
to the calibration mode.
CAL MODE
A CALIBRATING message appears for several seconds to indicate that the OCP
calibration is progressing and new OCP calibration constants of “SETUP 4” are
stored. Then the display will return to the calibration mode.
21
Turn off the power supply to exit the calibration mode.