Performance Tests
2
Keysight N1911A/1912A Service Guide
29
Test method
1
Enable the path that routes the time base signal to the trigger output
connector.
2
Using the 53132A, measure the frequency of the signal at the trigger output
connector.
NOTE
– This test can be configured manually via the command
SERV:BIST:TBAS:STAT ON
, which enables the 10 MHz feed to the trigger
output connector (refer to the programming guide for further details on the
use of this command).
– This test can also be configured manually via the front panel; access the
Service menu, select Self Test, and select Time Base to enable the 10 MHz
feed to the trigger output connector.
– No adjustment is available for this test if it fails (see
Chapter 4,
"Troubleshooting Guide"
).