5
Repair Guide
92
Keysight N1911A/1912A Service Guide
– Remove the rear panel (
Figure 5-14
): Use the
N1912-61807 special tooling kit to remove the
fasteners on the trigger connectors. Use the
9/32” socket to remove the GPIB standoffs.
Carefully pull the rear panel away from the main
board.
Figure 5-14
Instructions
Visual