Troubleshooting Guide
4
Keysight N1911A/1912A Service Guide
55
Performance Test
Type of failures
Debug tips
Possible faults
1 mW power reference level
failures
Attempt to adjust the 1 mW
power reference level
Calibrator assembly (high
probability)
Main board (low probability)
VSWR failures
Not applicable
Calibrator assembly
Time base frequency accuracy
failures
Check that the BNC cable being
used is not damaged
Check that the BNC is connected
to ‘Trig Out’, not ‘Trig In’
Main board
Zero set (average Path) failures
Not applicable
Main board
Zero set (peak path) failures
Not applicable
Main board
Linearity (average path) failures
Not applicable
Main board
Linearity (peak path) failures
Replace the DAP assembly for the
channel to see if this clears the
fault
DAP assembly (low probability)
main board (high probability)
Rise/fall time (peak path) failures Check/reseat the sensor flex RF
connections
Sensor flex assembly (low
probability)
DAP assembly (low probability)
Main board (high probability)