Replacing the PPMC Assembly
– The PPMC assembly is pre-programmed with N1912A firmware
– Always perform a firmware upgrade to the instrument if the PPMC assembly
has been replaced
– Instrument serial number:
This can be stored in the PPMC assembly via the command:
SERV:SNUM <CHARACTER DATA>
– Instrument option(s):
This/these can be stored in the PPMC assembly via the command:
SERV:OPT “<CHARACTER DATA>”
Refer to the
Programming Guide
for further details on the use of these commands.
NOTE
– Fitting a PPMC assembly that has been pre-programmed with N1912A
firmware to an N1911A power meter will generate errors; these errors will
disappear once the firmware upgrade procedure has been carried out.