Repair Guide
5
Keysight N1911A/1912A Service Guide
109
Additional Repair Notes
Replacing a sensor flex assembly:
– The sensor flex assembly is supplied straight
– Create a sharp bend (
Figure 5-20
): The flex circuit must be bent at a
right-angle where it meets the printed circuit board. It can only be bent after
heat has been applied to it (i.e. using a hot-air gun, or a similar device)
Figure 5-20
Creating a sharp bend
NOTE
– Once this sharp bend has been created, the flex should not be bent at this
point again; to do so may break the tracking within the flex.
– Route and connect the sensor flex assembly: once the sensor flex assembly
has been attached to the power meter; it should be folded to match the route
taken by the assembly being replaced. Heat may be used to assist the folding
of the flex.