Linearity (Average Path)
Description
Linearity over the full input voltage range of the measurement path iswarranted to
provide assurance of instrument accuracy. This test measures linearity by using a
calibration DAC and a calibration ADC (built into the DUT) to stimulate and
compare performance of the average path against the measurement ADC,
returning the worst case percentage error.
Equipment
– No test equipment required
Test method
1
Execute the internal linearity measurement procedure for channel A.
2
Read back the result of the measurement from the DUT.
3
If the DUT model number is N1912A, then repeat this procedure for channel B.
4
The test will take a few minutes to complete.
5
The measurement result should be less than 0.5 and greater than –0.5. The
optimum measurement result for this test is 0.
NOTE
– This test can be performed manually via the commands:
SERV:BIST:CW[1|2]:LIN
SERV:BIST:CW[1|2]:LIN:PERR?
(Refer to the programming guide for further details on the use of these
commands)
– No adjustment is available for this test if it fails (see
Chapter 4,
"Troubleshooting Guide"
).