5
Repair Guide
106
Keysight N1911A/1912A Service Guide
Step 6:
– Attach the inverter board to the display
interface board using the 2 screws removed
earlier
Step 7:
– Connect the inverter board to the display
interface board using the backlight cable
assembly
Note:
– The cable must be tucked under the plastic
clip to prevent any fouling
Step 8:
– Attach the calibrator assembly to the display
support molding using the 3 screws removed
earlier
Step 9:
– Carefully spread the EMI fingers outwards,
ensuring they extend beyond the edges of the
hole in which the calibrator assembly is fitted
Instructions
Visual