2
Performance Tests
34
Keysight N1911A/1912A Service Guide
Figure 2-6
Absolute accuracy test setup
Test Method
1
Connect the equipment as shown in
Figure 2-6
.
2
Unplug the power cord from the range calibrator. Eliminate ground loops to
enable the 11683A (Option H01) range calibrator to operate properly.
11683A range
calibrator front panel
11683A range
calibrator rear panel
BNC T-joint connector
BNC cable
BNC cable
N1911/1912A
P-Series power meter
3458A digital
multimeter
33250A function
generator
Banana
connector
Banana
connector
NOTE
The procedure details the key presses required on the Keysight N1911A P-Series
power meter. For the Keysight N1912A P-Series power meter, the equivalent key
presses should be performed on both channels.