34
6000 X-Series Oscilloscopes Service Guide
2
Testing Performance
To verify dual cursor accuracy
This test verifies the dual cursor accuracy for each analog channel.
This test is similar to the test for verifying the DC vertical gain, except you will
measure the dc voltage output of an precision source using dual cursors on the
oscilloscope and compare the results with the multimeter reading.
Dual cursor accuracy test limits: ±[DC vertical gain ac 0.42% full scale]
For the DC vertical gain accuracy test limits, see
1
Press
[Save/Recall] > Default/Erase > Factory Default
to recall the factory default
setup.
2
Set up the oscilloscope.
a
Set the Volts/Div setting to the value in the first line in
(depending
on the oscilloscope model).
b
Adjust the channel 1 position knob to place the baseline at 0.5 major division
from the bottom of the display.
Table 9
Equipment Required to Verify Dual Cursor Accuracy
Equipment
Critical Specifications
Recommended Model/Part
Precision Source
DC voltage of -5.5 V to 35.5 V, 0.1 mV resolution
Keysight B2912A/B2962A
Digital multimeter
Better than 0.01% accuracy
Keysight 34401A/34461A
50
Ω
BNC Cable (qty 2)
BNC - BNC, 48” length
Keysight 8120-1840
Adapter (qty 2)
BNC (f) to banana (m)
Keysight 1251-2277
BNC Tee
BNC tee (m) (f) (f)
Keysight 1250-0781 or Pomona 3285
Shorting cap
BNC
Keysight 1250-0774
Blocking capacitor
Note: if a BNC blocking capacitor is not available use
an SMA blocking capacitor.
Keysight
Pomona 4289 +
Pomona 5088
Summary of Contents for InfiniiVision 6000 X Series
Page 1: ...Keysight InfiniiVision 6000 X Series Oscilloscopes Service Guide...
Page 12: ...12 6000 X Series Oscilloscopes Service Guide Tables...
Page 58: ...58 6000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 82: ...82 6000 X Series Oscilloscopes Service Guide 4 Troubleshooting...