Replacing Assemblies
5
6000 X-Series Oscilloscopes Service Guide
93
To remove the AC input shield
1
To remove AC input shield, locate and remove using a TORX T10 driver the two
screws securing the shield to the rear deck.
2
Once the screws have been removed, carefully remove the shield by lifting it up
and off retaining tabs on rear deck.
Figure 26
Removing the AC input shield
WARNING
Thin sheet metal parts may have sharp edges. Handle with care to avoid injury.