74
6000 X-Series Oscilloscopes Service Guide
4
Troubleshooting
Figure 13
Verify power supply DC output
J102
8-10
3-5
Table 19
DC output from power supply
Supply
Test Point
Expected Value
If Good
If Bad
VP13V
Connector J102:
pins 8, 9, 10 = +13V
pins 3, 4, 5 = GND
+13 V ±0.2V
Indicates power supply is good.
Indicates power supply is bad or
cable from power supply to power
board connector J102 is bad.