Testing Performance
2
6000 X-Series Oscilloscopes Service Guide
37
8
Press the
Y2
softkey, then position the Y2 cursor to the center of the voltage
trace using the Entry knob.
The
Δ
Y value on the lower line of the display should be within the test limits of
If a result is not within the test limits, go to the “Troubleshooting” chapter. Then
return here.
9
Disconnect the precision source from the oscilloscope.
10
Repeat this procedure to check the dual cursor accuracy with the remaining
Volts/div setting values in
.
11
Finally, repeat this procedure for the remaining channels to be tested.
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 1 mV/div, 2 mV/div, and 5 mV/div, noise
may be a factor. To eliminate the noise, add a BNC Tee, blocking capacitor, and
shorting cap at the oscilloscope channel input to shunt the noise to ground. See
. If a BNC capacitor is not available, use an SMA blocking capacitor,
adapter, and cap. See
in the equipment list on
for
details.
Figure 6
Using a Blocking Capacitor to Reduce Noise
To oscilloscope input
BNC shorting
cap
Blocking
Capacitor
Summary of Contents for InfiniiVision 6000 X Series
Page 1: ...Keysight InfiniiVision 6000 X Series Oscilloscopes Service Guide...
Page 12: ...12 6000 X Series Oscilloscopes Service Guide Tables...
Page 58: ...58 6000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 82: ...82 6000 X Series Oscilloscopes Service Guide 4 Troubleshooting...