Troubleshooting
4
6000 X-Series Oscilloscopes Service Guide
81
Table 22
Power supplies to front panel board (AutoProbe power, from power board J106)
Supply
Test Point
Expected Value
If Good
If Bad
VP13VF_PRB
J106, pin 1
+13 V ±2%
Supply is good.
Replace the power board.
VM13VF_PRB
J106, pin 2
-13 V ±2%
Supply is good.
Replace the power board.
VM6V5_PRB
J106, pin 3
-6.512 ±3%
Supply is good.
Replace the power board.
VP6V5_PRB
J106, pin 4
+6.512 ±3%
Supply is good.
Replace the power board.