6000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Figure 3
Setting up Equipment for DC Vertical Gain Accuracy Test
6
Adjust the output so that the multimeter reading displays the first Volts/div
precision source setting value in
(depending on the oscilloscope
model).
7
Disconnect the multimeter.
8
Wait until the measurement settles.
9
Read the “current” average voltage value again as V2.
10
Calculate the difference V2 - V1.
Precision Source
Oscilloscope
Digital
Multimeter
BNC Tee
BNC (f) to dual
bananna adapter