66
6000 X-Series Oscilloscopes Service Guide
4
Troubleshooting
Figure 10
Default setup screen
2
If your screen looks substantially different, replace the acquisition board.
To perform an Auto Scale on the Probe Comp signal
1
Press the
[Default Setup]
key. The oscilloscope is now configured to its default
settings.
2
Connect an oscilloscope probe from channel 1 to the
Probe Comp
signal terminal
on the front panel.
3
Connect the probe’s ground lead to the ground terminal that is next to the
Demo
2
(Probe Comp) terminal.
4
Press
[Auto Scale]
.