Testing Performance
2
6000 X-Series Oscilloscopes Service Guide
29
To verify DC vertical gain accuracy
This test verifies the accuracy of the analog channel DC vertical gain for each
channel.
In this test, you will measure the dc voltage output of a precision source using the
oscilloscope’s
Average - Full Screen
voltage measurement and compare the results
with the multimeter reading.
1
Press
[Save/Recall] > Default/Erase > Factory Default
to recall the factory default
setup.
Table 6
DC Vertical Gain Accuracy Test Limits
Models
Test Limits
Notes
6000 X-Series
±2.5% of full scale
•
Full scale is defined as 16 mV on the 1 mV/div range.
•
Full scale on all other ranges is defined as 8 divisions
times the V/div setting.
Table 7
Equipment Required to Verify DC Vertical Gain Accuracy
Equipment
Critical Specifications
Recommended Model/Part
Precision Source
DC voltage of -5.5 V to 35.5 V, 0.1 mV resolution
Keysight B2912A/B2962A
Digital multimeter
Better than 0.01% accuracy
Keysight 34401A/34461A
50
Ω
BNC Cable (qty 2)
BNC - BNC, 48” length
Keysight 8120-1840
Adapter (qty 2)
BNC (f) to banana (m)
Keysight 1251-2277
BNC Tee
BNC tee (m) (f) (f)
Keysight 1250-0781 or Pomona 3285
Shorting cap
BNC
Keysight 1250-0774
Blocking capacitor
Note: if a BNC blocking capacitor is not available use
an SMA blocking capacitor.
Keysight
Pomona 4289 +
Pomona 5088
Summary of Contents for InfiniiVision 6000 X Series
Page 1: ...Keysight InfiniiVision 6000 X Series Oscilloscopes Service Guide...
Page 12: ...12 6000 X Series Oscilloscopes Service Guide Tables...
Page 58: ...58 6000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 82: ...82 6000 X Series Oscilloscopes Service Guide 4 Troubleshooting...