Getting Started Guide 99
Internal Application Server Operating System
System Maintenance
— Any upgrades that were made to the Keysight Test Application software.
It is recommended that routine back-ups of the instrument information be
performed to keep current archives of the instrument information. This allows a
full recovery of the instrument information after the instrument recovery
system operations are performed. See
for more details.
Using the Instrument Image Recovery System - Internal Application
Server Operating System
It is recommended that you use a regular back-up strategy. Your IT department
may already have a back-up strategy in place which is suitable for the Internal
Applications Server and its data. See
. Using the Instrument Image Recovery System in conjunction with a
regular back-up strategy should allow you to fully recover the Internal
Applications Server software and data.
Step
Notes
1.
Check the size of
D:\Users\
Administrator
and
D:\Users\
Instrument
.
The size must be
smaller than the half
the free space in the
D:
drive.
If the used space is bigger than half the free space in the D: drive, proceed as follows:
(a) Remove the contents of
D:\Users\Administrator\AppData\Local\Temp
(b) Move the contents of Desktop, Documents, Images, and Music to
D:\Administrator
(c) Verify that the size of
D:\Users\Administrator
is smaller now.
(d) Repeat the process for
D:\Users\Instrument
, moving the folders into
D:\Instrument
.
2.
Turn the instrument
off and make sure
there are no Ethernet
cables connected to
the instrument.
3.
Connect an external
mouse, keyboard, and
display to the
Internal Application
Server.