8163A/B, 8164A/B, 8166A/B User’s Guide
447
Cleaning Instructions
15
How to clean optical devices which are sensitive to mechanical stress and pressure
Some optical devices, such as the Keysight 81000BR Reference Reflector,
which has a gold plated surface, are very sensitive to mechanical stress or
pressure. Do not use cotton-swabs, soft-tissues or other mechanical
cleaning tools, as these can scratch or destroy the surface.
Preferred Procedure
Use the following procedure on most occasions.
1 Blow away any dust or dirt with compressed air.
Procedure for Stubborn Dirt
To clean devices that are extremely sensitive to mechanical stress or
pressure you can also use an optical clean polymer film. This procedure is
time-consuming, but you avoid scratching or destroying the surface.
1 Put the film on the surface and wait at least 30 minutes to make sure
that the film has had enough time to dry.
2 Remove the film and any dirt with special adhesive tapes.
Alternative Procedure
For these types of optical devices you can often use an ultrasonic bath
with isopropyl alcohol. Only use the ultrasonic bath if you are sure that it
won't cause any damage anything to the device.
1 Put the device into the bath for at least three minutes.
2 Blow away any remaining liquid with compressed air.
If there are any streaks or drying stains on the surface, repeat the cleaning
procedure.
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......