236
8163A/B, 8164A/B, 8166A/B User’s Guide
8
Setting Attenuation and/or Power Levels
Attenuator Status Indicators
Status Indicator
Description
B (or Busy)
The Attenuator’s filter is transitioning to a new setting.
Offs
Any combination of an
a
offset, P offset, or Offset (
λ
)
has been applied.
Exa
An exact
λ
offset has been applied.
Int
An interpolated
λ
offset has been applied.
Ext
An extrapolated
λ
offset has been applied.
PC
The Power Control feature is on (81576A/77A)
Exp
(81566A/67A/76A/ 77A
only)
There is either too much, or too little, input power for the
Attenuator module to set or maintain your desired P
SET
.
RiL
The Remote Interlock Connector of your mainframe is open. No
measurement is possible until the connection at the
Remote Interlock Connector closes.