122
8163A/B, 8164A/B, 8166A/B User’s Guide
3
Power Measurement
•
<
Averaging Time Over
>, a trigger is output when the averaging time
period of a measurement finishes.
•
<
Meas Started
>, a trigger is output when the averaging time period
of a measurement begins.
•
<Disabled
>, the output trigger mode is disabled.
5 Press
Enter
.
Dual Power Meters - Master and Slave Channels
For the Keysight 81635A Dual Power Sensor and the Keysight 81619A
Dual Optical Head Interface module, channel 1 is the master channel and
channel 2 is the slave channel. The master and slave channels share the
same software and hardware triggering system. For some commands,
setting parameters for the master channel sets the parameters for the
slave channel. In these cases, you may only set parameters for the slave
channel by setting master channel parameters.
The parameters listed in
on page -122 can only be set or viewed
using the master channel.
Table 12
Parameters that can only be set using the master channel
Parameter
<
Zero
>
<
MinMax mode
>
<
Reset MinMax
>
<
Data points
>
<
Averaging time
>
<
Range mode
>
<
Input trigger mode
>
<
Output trigger mode
>
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......