180
8163A/B, 8164A/B, 8166A/B User’s Guide
7
Return Loss Measurement
How to Choose a Light Source
Highly accurate return loss measurement requires that you use a light
source with a subset of the following properties:
• low coherence length
• high power
• [and, for the 81534A, power stability.]
If a return loss module with at least one internal source (such as the
81611A, 81612A or 81613A) is not available, we recommend that you use
any of the following as part of your return loss measurement setup:
•
Erbium ASE Source.
This source offers a high-power low-coherence output that is very
stable over time.
•
Laser Source modules
.
on page 345 for a list of modules that can be installed
in your mainframe. Refer to the note below that explains the use of
highly coherent light sources.
•
Tunable Lasers,
when you want to measure Return Loss over a
wavelength range.
on page 345 for a list of modules that can be installed
in your mainframe. Refer to the note below that explains the use of
highly coherent light sources.
•
LED sources
.
The intensity of the output of LED sources is very stable over time,
although the low power output of LEDs restricts the dynamic range of
return loss measurement.
NOTE
If you use high-coherence light sources, you can improve performance
by:
•
modulating the output signal at 2 kHz or higher, or
•
using coherence control to reduce the coherence length of the signal.
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......