8163A/B, 8164A/B, 8166A/B User’s Guide
341
Installation and Maintenance
11
Serial Interface
The mainframe’s serial interface has configurable and fixed parameters.
• The Baudrate set for the serial interface of the instrument should match
that set for the PC serial interface connected to it.
• The PC serial interface should be configured to match the instrument’s
fixed parameters.
Configurable Parameters
Baudrate. Refer to
How to Set the Speed of the Serial Interface
Fixed Parameters
These are:
Serial Interface Configuration - Fixed Parameters
Data Bits
8
Parity
None
Stop Bits
1
Flow Control
Hardware (RTS/CTS)
NOTE
For serial communication use the null modem cable provided with your
instrument (Keysight P/No. 5180-2477). This is primarily used for the
firmware upgrade process.