190
8163A/B, 8164A/B, 8166A/B User’s Guide
7
Return Loss Measurement
Tip:
The Keysight 81000BR Back Reflector is such a component, offering a
return loss of 0.18
±
0.1 dB.
3 Move to the Return Loss channel, then press [Details]
4 Move to [
FPDelta
], press [Edit]
,
set the value to 0.000, then press [OK].
5 Move to [
RLref
] and make sure that the displayed value is correct. Set
[
RLref
] to the value of the return loss of the reference reflection you are
using. For example, if you are using the Keysight 81000BR reference
reflector, set [
RLref
] to 0.18 dB.
Figure 109
Measuring the Reflection Reference
6 Switch on the Laser.
7 Press [Menu]. Move to <
Reflectance calibration
> and press
Enter
. The
instrument measures the power reflected by the back reflector. The
[
RL
] value changes to the same value as entered for [
RLref
].
Tip:
An alternative step
is to select the Return Loss module’s
[Detail] screen, then press [RefCal]