8163A/B, 8164A/B, 8166A/B User’s Guide
389
Error Messages
14
Selftests
These tests are started after initialization of the instrument. This test
measures defined parameters, and tests them to limits.
error message
Initialization Error - module temperature too high
(please wait a little bit and try again)!
description
Module temperature is too high
error value
(no error value)
action
Doublecheck whether the fan is working and not blocked. Wait some time and reboot the
instrument. If the failure persists, please return the unit to your local service center for repair.
error message
Main Peltier Temp. Selftest Failed (error value)
description
Verifies whether the peltier temperature of the laser diode is within a given range
error value
13010
Peltier temperature is not in target range
action
Please return the unit to your local service center for repair
error message
Main Thermo Temp. Selftest Failed (error value)
description
Verifies whether the temperature of the thermopositioner is within a given range
error value
13020
Thermopositioner is not in target temperature range
action
Please return the unit to your local service center for repair
error message
Main Peltier Curr. Selftest Failed (error value)
description
Checks if peltier current for the laser diode peltier is below maximum
error value
13050
Laser peltier current is above maximum value
action
Please return the unit to your local service center for repair
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......