SAFETY RULES

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Maximum Overload Protection Input

 

 

 

DESCRIPTION OF PANEL

 

Protection cover

Test leads store space

Display

Power switch

and

Function switch

Test  lead  (Black)

for Negative  (- )

Select switch

Auto hold

Test  lead  (Red)

for Positive (+)

 

 
 

 

DESCRIPTION OF FUNCTIONS

 

 

Power Switch and Function Switch

 

    Turn the switch for power on and off and to select the functions 

of DCV, ACV, 

, Buz, Diode. 

 

Battery Voltage Drop Indication Display 

    If the internal battery has been consumed and the voltage drops, 

the display shows. 

  In this case, replace with 2 new batteries.(LR-44 or SR-44). 

 

Auto Hold Function 

  Press A-HOLD key. (The display shows the symbol)

 

  Connect the test leads to the object under test.

 

  When the reading stabilizes, the buzzer sounds. 
  Remove the testing leads from the object under test. 
  The D.M.M now shows the measured value that it retains. 
  You can repeat steps   to   as many times as you like as 

long as the display shows the symbol. 

  To cancel this function, press the A-HOLD key once again. 

Note: 
Do not mistake the following for a malfunction! 

 In DC voltage measurement, the Auto Hold function is only 

available for range over 4V. 

 The Auto Hold function cannot be applied to unstable 

signals. 

 

 

MEASUREMENT RANGES

 

 23±5

°

C <80%RH.      Accuracy: ± (% r digit) 

V

 

Range Resolution Accuracy

Input 

Resistance 

Maximum 

input 

400mV 0.1mV 

1.2+2 

>100M

 Ω

 

4V 0.001V  0.7+1 11M

 Ω

 

40V 0.01V 
400V 0.1V 
600V 1V 

 
1.2+1 

 
10M

 Ω

 

 
 

600V 

V

 

Range Resolution Accuracy

Input 

Resistance 

Maximum 

input 

4V 0.001V 

11M

, <50pF 

40V 0.01V 
400V 0.1V 
600V 1V 

 
2+5 

 
10M

, <50pF 

600Vrms 

Resistance 

Range Resolution Accuracy

Measuring cur-

rent and Open 

voltage 

Maximum 

input 

400

 Ω

 0.1

 Ω

 

<1mA, <3.4V 

4k

 Ω

 0.001k

 Ω

 

<0.5mA, <1.0V

40k

 Ω

 0.01k

 Ω

 <70

µ

A, <0.7V 

400k

 Ω

 0.1k

 Ω

 

 
1.2+2 

<7

µ

A, <0.7V 

4M

 Ω

 0.001M

 Ω

 

2+3 

<0.7

µ

A, <0.7V 

40M

 Ω

 0.01M

 Ω

 

5+3 <70nA, 

<0.7V 

 
 

600V 

 

Range Resolution Accuracy

Open-Circuit 

Voltage 

Maximum 

input 

400

 Ω

 0.1

 Ω

 

Beep on

50±20

 

<3.4V 600V 

 

Range Resolution Accuracy

Open Circuit 

Voltage 

Maximum 

input 

2V 0.01V  1.5+1 <3.4V 

600V 

 
 
 
 
 

 

GENERAL SPECIFICATIONS

 

Measuring method 

:  Dual integration mode 

Display 

:  Counter approx. 4300 counts max. 

Range selection 

:  Auto range 

Over indication 

:  “OL” mark.(Voltage only) 

“---” mark. (Others) 

Polarity 

: Automatic selection (“-” is displayed 

only.) 

Low battery indication  :  If the internal battery has been   

consumed and the voltage drops, the 
display shows mark. 

Sampling rate 

:  Approx. 2 times/sec. 

Operating tenperature/ :  0~50°C <80%RH.   
Humidity  

No 

condensation 

Storage temperature/  :  -10~60°C <70%RH. 
Humidity  

No 

condensation 

Power supply 

:  LR-44 

× 

2 or SR-44 

× 

Dimension & weight 

:  177 (L) 

× 

76 (W) 

× 

18 (H) mm 

Approx. 110g 

Accessories 

:  Instruction Manual ………………1 

LR-44 Button Cell (installed)……2 

Compliance with Standards 
 Safety 

EN61010-1 

 

 

 

(600V CAT I, 300V CAT II) 

 EMC 

EN61326 

 

   

EN55022  (Class B) 

 

 

MEASUREMENT PROCEDURE

 

 
 
 
 
 
 
 
 
 

 

DC Voltage (DCV) 

  DCV Measurement Maximum Rating Input Value 
    600V DC (CAT. I) 

1) Applications. 
 Measure 

D.C. 

circuits. 

2) Measuring Ranges. 
 

5 ranges from 400mV to 600V. 

3) Measurement Procedure. 

  Set the function switch at V (DCV) range. 
  Apply the black test pin to the negative potential side of 

the circuit to measure and the red test pin to the positive 
potential side. 

  Read the value on the display. 
  After measurement, remove the red and black test pins 

from the circuit measured. 

  The display fluctuates when the test leads are removed. 

This is not malfunction. 

 

AC Voltage (ACV) 

    ACV Measurement Maximum Rating Input Value 
  600Vrms (CAT. I) 

1) Applications. 

Measure sine-wave A.C. voltages such as lighting voltages. 

2) Measuring Ranges 

4 ranges from 4V to 600V. 

3) Measurement Procedure 

 

Set the function switch at ACV range.

 

 

Apply the red and black test pins to the circuit to   
measure. 

 

Read the value on the display. 

 

After measurement, remove the red and black test pins 
from the circuit measured.

 

  This instrument employs the average measurement   

system and some error is made to the display of   
waveforms other than sine waves.

 

 The accuracy guaranteed frequency range is 45 Hz to 400 

Hz.

 

 

Resistance Measurement 

 
     
 

1) Applications 

Resistance of resistors and circuits is measured. 

2) Measuring Ranges 

6 ranges from 400

 Ω

 to 40M

 Ω

3) Measurement Procedure. 

  Set the function switch of 73101 at OHM. 
  Apply the red and black test pins to an object to   

measure. 

  Read the value on the display. 
  After measurement, release the red and black test pins 

from the object measured. 

  If measurement is likely to be influenced by noise, shield the 

object to measure with negative potential(“-” test lead black). 

  If a test pin is touched by a finger during measurement, 

measurement will be influenced by the resistance in the 
human body to result in measurement error. 

 

Checking Continuity (

 
 
 

1) Applications 

Checking the continuity of wiring and selecting wires. 

2)  How to Use 

  Set the function switch of 73101 at 

 Apply the red and black test pins to a circuit or 

conductor to measure. 

  The continuity can be judged by whether the buzzer 

sounds or not. 

 After measurement, release the red and black test pins 

from the object measured. 

 The buzzer sounds when the resistance in a circuit to 

measure is less than about 70

 Ω

 

Testing Diode (

)

 

 
 
 

1) Applications 

The quality of diodes tested. 

2)  How to Use 

  Set the function switch at 

 range. 

  Connect the test leads to the diode to be tested. 

When measuring the forward voltage across diode,   
a normal diode will indicate 0.5 to 0.7V . 
The reverse voltage will indicate “OL”. 

  For a short-circuited diode, a value near 0V will be   

displayed. 

 

 

AUTO POWER OFF SELECTION

 

  The D.M.M will power off automatically in 20 minutes later 

after the last operation was stopped. 

  The alarm buzzer sounds 7 times before around 1 minute of 

power off. 

  Press any key or turn the switch can postpone the power-off 

time or restart the D.M.M is turned off. 

  To cancel the Auto Power Off Function, hold down the 

SELECT key and then set the function switch from OFF to 
the position of any desired measurement mode. 

  The AUTO POWER OFF indication turns off when the 

Function is cancelled. 

 

 

MAINTENANCE

 

 

Battery Replacement 

 
 
 
 
 
 
 
 
 
 

(How to Replace) 
     Remove the battery lid screw with a screwdriver. 
     Remove the battery cover. 
     Take out the batteries and replace with 2 new batteries. 
     Attach the battery lid and screw it on. 

 
 
 
 

 

YOKOGAWA M&C CORPORATION 
International Sales Dept. 

2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan 
Phone: 81-422-52-5716  Facsimile: 81-422-55-8654 

YOKOGAWA CORPORATION OF AMERICA (U.S.A.) 

Phone: 1-770-253-7000  Facsimile: 1-770-251-2088 

YOKOGAWA EUROPE B. V. (THE NETHERLANDS) 

Phone: 31-334-64-1611    Facsimile: 31-334-64-1610 

YOKOGAWA AMERICA DO SUL S. A. (BRAZIL) 

Phone: 55-11-5681-2400  Facsimile: 55-11-5681-1274 

YOKOGAWA ENGINEERING INSTRUMENTS KOREA CORPORATION (KOREA) 

Phone: 82-2-551-0660 to -0664    Facsimile: 82-2-551-0665 

YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA) 

Phone: 61-2-9805-0699  Facsimile: 61-2-9888-1844 

YOKOGAWA BLUE STAR LTD. (INDIA) 

Phone: 91-80-227-1513  Facsimile: 91-80-227-4270 

LTD. YOKOGAWA ELECTRIC (RUSSIAN FEDERATION) 

Phone: 7-095-737-7868  Facsimile: 7-095-737-7869 

 

KIM3E-2003.2 

Function Maximum 

rating 

input value 

Maximum over- 

Load protection 

input 

V (DCV) 

DC 600V 

V (ACV) 

AC 600V 

 
 
 

Voltage and current 
input prohibited. 

 
DC 600V 
 
AC 600V rms 

User’s 

Manual 

73101 

Pocket Digital Multimeter

 

Please read this instruction manual thoroughly in 
order to use this pocket DMM safely. And for 
reference, please keep this manual together with the 
meter. 
Instructions followed the sign “WARNING” must be 
obeyed to avoid accidental burn or electric shock. 

To avoid damage to instrument or electrical shock! 
The maximum input voltage level for 73101 depend 
on the over-voltage categories specified by the safely 
standards. 
These categories are described as below to protect 
operators against transient impulse voltages in power 
lines. 

Over-Voltage 
Category (CAT.) 

Maximum Input 

Voltage 

I 600V 

II 300V 

Over-voltage category I (CAT. I): 
Covers the path from a main outlet, through a power 
transformer, to a circuit in the transformer’s secondary 
stage. 
Over-voltage category II (CAT. II): 
Covers a primary-stage circuit of equipment 
connected to a main outlet. 

To make sure that the meter is used safely, the owner 
has to follow the instructions while using the 
instrument. 
1.  Pay special attention when measuring the 

voltage of AC 30Vrms or DC 60V or more to 
avoid injury. 

2.  Never apply an input signals exceeding the 

maximum rating input value. 

3.  Never use meter for measuring the line 

connected with equipment (i.e. motors) that 
generates induced or surge voltage since it may 
exceed the maximum allowable voltage. 

4.  Never use meter if the meter or test leads are 

damaged or broken. 

5.  Never use uncased meter. 
6.  When connecting the test leads, first connect the 

ground lead(black). When disconnecting them, 
the ground lead must be disconnected last. 

7.  Always keep your fingers behind the finger 

guards on the probe when making 
measurement. 

8.  Be sure to disconnect the test pins from the 

circuit when changing the function. 

9.  Before starting measurement, make sure that the 

function is properly set in accordance with the 
measurement. 

10.  Never use meter with wet hands or in a damp 

environment. 

11.  Never open tester case except when replacing 

batteries. Do not attempt any alteration of original 
specifications. 

1.  Never apply an input signals exceeding the maximum 

rating input value. 

2.  Be sure to disconnect the test pins from the circuit 

when changing the function. 

3.  Always keep your fingers behind the finger guards on 

the probe when making measurement. 

  Never apply voltage to the input terminals. 

  Never apply voltage to the input terminals. 

  Never apply voltage to the input terminals. 

1. If the rear case or the battery lid is removed with 

input applied to the input terminals, you may get 
electric shock. Before starting the work, always 
make sure that no input is applied. 

2. Before starting the work, be sure to turn OFF the 

main unit power and release the test lead from the 
circuit. 

WARNING

 

WARNING

 

WARNING

 

WARNING

 

WARNING

 

WARNING

 

WARNING

 

IM 73101 

1st Edition: 2003. 4 