8163A/B, 8164A/B, 8166A/B User’s Guide
301
Applications
10
On-Screen Messages
The messages listed in the table below may appear at the top of the screen
during the PACT application.
Table 18
PACT Application On-Screen Messages
Keysight 8163A/B
Message
Keysight 8164A/B
Message
Description
SLck
SLock
The Tunable Laser module is locked. No measurement is
possible until you unlock the module, see
Lock/Unlock the High-Power Laser
on page 67 for details.
RiLck
RILock
The Remote Interlock Connector of your mainframe is
open. No measurement is possible until the connection
at the Remote Interlock Connector closes, see
Remote Interlock (RIL) connector
page 337 for details.
NoH
No Head
There is an Optical Head Interface module inserted in the
mainframe that does not have an Optical Head
connected to it. No measurement is possible using this
channel until you either:
reconnect an Optical Head or
deselect this Optical Head channel as a power
measurement channel.
NoD
No Data
No data has been measured or measurement is not yet
finished. You cannot access the Analysis Screen yet.
Printout and file operations are not yet possible.
Set
Settling
A Tunable Laser module is settling. The instrument is
busy and no measurement can be performed yet.
NoR
No Ref
No reference sweep has been performed. Loss will be
calculated using nominal laser power as a reference.
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......