8163A/B, 8164A/B, 8166A/B User’s Guide
14
Error Messages
Tunable Laser Sources
Initialization Tests
These tests are executed every time the instrument is powered on, before
the selftests start.
error message
Initialization Error (AttInitialize(), error value)
description
Error during attenuator initialization. No ready message from attenuator DSP after ...
error value
1
reset
2
init motor
4
find index
8
move to min position
action
Please return the unit to your local service center for repair
error message
Initialization Error (AttMovePos(zeroPos), error value)
description
No ready message from attenuator DSP after moving to zero position taken from caldata
error value
8
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (CalData CRC wrong:
crc1
, stored:
crc2
!)
description
The checksum stored in the caldata does not match the calculated checksum
error value
crc1
calculated checksum
crc2
checksum read from caldata
action
Please return the unit to your local service center for repair