Calibrating Probes
3
1130B-Series Probes User’s Guide
81
Step 2. Verifying the Connection
1
On the Infiniium oscilloscope, press the autoscale button on the front panel.
2
Set the volts per division to 100 mV/div.
3
Set the horizontal scale to 1.00 ns/div.
4
Set the horizontal position to approximately 3 ns. You should see a waveform
similar to that in
If you see a waveform similar to that of
, then you have a bad connection
and should check all of your probe connections.
Figure 49
Connecting the Probe and Deskew Fixture
BNC to SMA Connector
Deskew Fixture
50W SMA Terminator
InfiniiMax Probe
Pincher