208
1130B-Series Probes User’s Guide
9
Performance Verification
Figure 226
PV Fixture Connected to VNA
11
Perform a calibration at the reference plane:
a
Press the
reflection
screen key.
b
Connect the open end of 85052D Calibration Standard
to the non-pincher
side of the PV/DS test board.
c
Select the
open
screen key under the
Forward
group.
d
Wait until the VNA beeps indicating that it has completed the task.
e
Connect short end of Calibration Standard to the non-pincher side of the
PV/DS test board.
f
Select
short
screen key under the
Forward
group.
g
Wait until the VNA beeps indicating that it has completed the task.
h
Connect load end of Calibration Standard to the non-pincher side of the
PV/DS test board.
i
Select the
loads
screen key under the
Forward
group.
j
Press
broadband
screen key selection.
k
Wait until the VNA beeps indicating that it has completed the task.
l
Press the
done loads
screen key.
m
You have just calibrated one side of the reference plane.
12
Connect the other high quality SMA cable to the VNA’s
PORT 2
connector.
Summary of Contents for 1130B Series
Page 1: ...Keysight 1130B Series Differential and Single Ended Probes User s Guide ...
Page 6: ...6 1130B Series Probes User s Guide ...
Page 59: ...Using Probe Heads 2 1130B Series Probes User s Guide 59 Figure 33 N5380B Schematic ...
Page 78: ...78 1130B Series Probes User s Guide 2 Using Probe Heads ...
Page 86: ...86 1130B Series Probes User s Guide 3 Calibrating Probes Figure 53 Overlapping Waveforms ...
Page 90: ...90 1130B Series Probes User s Guide 3 Calibrating Probes ...
Page 98: ...98 1130B Series Probes User s Guide 4 Characteristics and Specifications ...
Page 226: ...226 1130B Series Probes User s Guide 10 SPICE Models ...
Page 228: ...228 1130B Series Probes User s Guide 10 SPICE Models ...