Performance Verification
9
1130B-Series Probes User’s Guide
211
33
Connect probe tip under the PV fixture’s pincher. Apply upward pressure to the
clip to ensure a proper electrical connection. Place the probe’s "+" side on
center conductor and "-" side to ground as shown in the following figure.
Figure 229
Probe Locations on PV Fixture
34
Press the
[Sweep Setup]
key on the VNA. Then press the
trigger menu
screen key.
Select the
continuous
screen key.
35
The V
in
waveform shown on screen should be similar to that shown in
Figure 230
Typical V
in
Waveform for an 1134B Probe
36
Select
[display]
key then
data->memory
screen key.
37
You have now saved V
in
waveform into the VNA's memory for future use.