212
1130B-Series Probes User’s Guide
9
Performance Verification
Measuring Vout Response
38
Disconnect the
PORT 2
cable from PV/DS test board and attach to probe output
on the AutoProbe Adapter.
39
Connect the Calibration Standard load to PV/DS test board (non-pincher side).
40
Press
[scale reference]
key on the VNA.
41
Set reference value to –20 dB.
42
Hold probe in place as described previously.
43
The display on screen is V
out
and it should be similar to that shown in
Figure 231
Typical V
out
Waveform for an 1134B Probe
Displaying the Vout/Vin Response
44
Press the
[Display]
key.
45
Then select the
Data/Memory
screen key. You may need to adjust the
Reference
Value
, located under the
Scale Ref
key, slightly to position the waveform at
center screen. The waveform should be similar to that shown in
Summary of Contents for 1130B Series