Performance Verification
9
1130B-Series Probes User’s Guide
213
Figure 232
Typical Waveform for an 1134B Probe
46
Press marker key and position the marker to the first point that the signal is –3
dB below center screen.
47
Read marker frequency measurement and record it in the test record located
later in this chapter.
48
The bandwidth test passes if the frequency measurement is greater that the
probe's bandwidth limit.