Calibrating Probes
3
1130B-Series Probes User’s Guide
87
Calibration for Hand-held Browser Probe Heads
Calibration of the hand-held browser probe heads consists of a vertical calibration
and a skew calibration. The vertical calibration should be performed before the
skew calibration. Both calibrations should be performed for best probe
measurement performance.
Calibration Setup
The calibration procedure requires the following parts.
• BNC (male) to SMA (male) adapter
• Deskew fixture
• 50
Ω
SMA terminator
1
As shown in
on page 88, connect BNC (male) to SMA (male) adapter
to the deskew fixture on the connector closest to the yellow pincher.
2
Connect the 50
Ω
SMA terminator to the connector farthest from the yellow
pincher.
3
Connect the BNC side of the deskew fixture to the Aux Out of the Infiniium
oscilloscope.
4
Connect the probe to an oscilloscope channel.
5
Place the positive resistor tip of the browser on the center conductor of the
deskew fixture between the green line and front end of the yellow pincher. The
negative resistor tip or ground pin of the browser must be on either of the two
outside conductors (ground) of the deskew fixture.
6
On the Infiniium oscilloscope in the Setup menu, select the channel connected
to the probe.
7
In the Channel Setup dialog box select the Probes... button.
8
In the Probe Setup dialog box select the Calibrate Probe... button.
9
In the Probe Cal dialog box select the Calibrated Atten/Offset radio button.
10
Select the Start Atten/Offset Calibration... button and follow the on-screen
instructions for the vertical calibration procedure.
11
Once the vertical calibration has successfully completed, select the Calibrated
Skew... button.
12
Select the Start Skew Calibration... button and follow the on-screen
instructions for the skew calibration.
NOTE
Before calibrating the probe, verify that the Infiniium oscilloscope has been calibrated recently
and that the calibration
Δ
temperature is within ±5 °C. If this is not the case, calibrate the
oscilloscope before calibrating the probe. This information is found in Infiniium Calibration
dialog box.
Summary of Contents for 1130B Series
Page 1: ...Keysight 1130B Series Differential and Single Ended Probes User s Guide ...
Page 6: ...6 1130B Series Probes User s Guide ...
Page 59: ...Using Probe Heads 2 1130B Series Probes User s Guide 59 Figure 33 N5380B Schematic ...
Page 78: ...78 1130B Series Probes User s Guide 2 Using Probe Heads ...
Page 86: ...86 1130B Series Probes User s Guide 3 Calibrating Probes Figure 53 Overlapping Waveforms ...
Page 90: ...90 1130B Series Probes User s Guide 3 Calibrating Probes ...
Page 98: ...98 1130B Series Probes User s Guide 4 Characteristics and Specifications ...
Page 226: ...226 1130B Series Probes User s Guide 10 SPICE Models ...
Page 228: ...228 1130B Series Probes User s Guide 10 SPICE Models ...